Mounting the Novumetrics unit
- Place the base plate against a level wall with at least 2.5 m of vertical clearance.
- Secure anchor bolts using the provided template and torque wrench (35 Nm).
- Attach the vertical mast, ensuring the spirit level indicator stays centered.
- Snap the soft arm-span wings into their magnets; they auto-register during boot.
Tip: allow 0.5 m of clearance on each side for smooth arm span guidance.
Power and network
- 110–240 V AC input · 150 W peak draw · IEC C13 power cable (included).
- Ethernet preferred (802.3af PoE optional); Wi-Fi 6 supported for temporary deployments.
- Offline mode caches up to 500 sessions with later sync via USB-C export.
For VLAN or firewall allowlists, open outbound ports 443 and 123 to *.novumetrics.cloud.
First calibration routine
- Insert the calibration wand until it clicks; the UI confirms sensor alignment.
- Follow on-screen prompts to place the included 100 cm and 180 cm rods.
- Scan the provided QR code to upload calibration logs to the Novumetrics portal.
- Remove the wand; the system auto-locks to prevent drift.
Weekly micro-calibrations take 90 seconds and can be scheduled before clinics open.
Training your staff
- Provide the 15-minute onboarding video included in the portal.
- Use the demo mode to rehearse cues with colleagues—no data is stored.
- Create at least two operator accounts to maintain audit trails.
Need custom onboarding? Email success@novumetrics.nl for remote workshops.
Printable launch checklist
Post this in your equipment room so every operator can validate setup in under two minutes.
Download checklist PDF